Monochromatic Ag X-ray Source - Page 3 |
|
|
|
|
Written by Adam Roberts
|
|
Thursday, 23 November 2006 15:30 |
|
Page 3 of 4
Results
Performance on a sputter cleaned Silver standard: >10000cps at 1.4eV FWHM >1000 cps at <0.9eV FWHM
Figure 2:Gold survey spectrum excited by Ag-mono
Gold foil (cleaned, survey spectrum: Fig. 2).Au 4f 7/2 @1.3eV FWHM 4,500cps (Edgell et al 25cps)
Greater analysis depth The Al 2p spectra shown (Figure 3) were acquired with (a) Al and (b) Ag sources. Spectrum (b) is a clear illustration of the greater analysis depth of the Ag source as the element : oxide ratio is greater than that acquired with the Al source. An oxide thickness of 1.8nm indicated here was confirmed by independent angle resolved measurements.
Figure 3: Al 2p regions:Al mono (a), and Ag mono (b) sources
|
|